RESOLUTION CLIP CLIPPING DEVICE M00522612 2261

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-20 for RESOLUTION CLIP CLIPPING DEVICE M00522612 2261 manufactured by Boston Scientific Corporation.

Event Text Entries

[188526834] Patient's exact age is unknown; however, it was reported that the patient was over the age of 18. (b)(4). Although the suspect device has been received, the evaluation has not been completed. Therefore, the cause of the reported malfunction has not been determined. Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[188526835] It was reported to boston scientific corporation that a resolution clip device was used in the ascending colon during a colonoscopy procedure performed on (b)(6) 2020. According to the complainant, during the procedure, the clip grasped and locked onto tissue, but failed to release from the catheter to deploy. Reportedly, during device removal, the clip got lodged in the colonoscope. It was also noted that the sheath was removed from the device prior to passing the clip through the scope. The procedure was completed with another resolution clip device. There were no patient complications reported as a result of this event. The patient condition at the conclusion of the procedure was reported to be fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2020-01249
MDR Report Key9862466
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-20
Date of Report2020-03-20
Date of Event2020-02-26
Date Mfgr Received2020-02-26
Device Manufacturer Date2019-02-21
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
Manufacturer Street2546 CALLE PRIMERA PROPARK, COYOL
Manufacturer CityALAJUELA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRESOLUTION CLIP CLIPPING DEVICE
Generic NameCLIP, HEMOSTATIC
Product CodeMCH
Date Received2020-03-20
Returned To Mfg2020-03-19
Model NumberM00522612
Catalog Number2261
Lot Number0023344081
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-20