NEOPUFF INFANT RESUSCITATOR RD900

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-20 for NEOPUFF INFANT RESUSCITATOR RD900 manufactured by Fisher & Paykel Healthcare Ltd.

Event Text Entries

[187414638] (b)(4). Method: the complaint rd900 neopuff infant resuscitator was returned to fisher & paykel healthcare (f&p) regional office in (b)(4) and was inspected by a trained f&p technician. Result: visual inspection of the returned neopuff revealed the gas outlet and inlet port were broken. Conclusion: the neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to infants until adequate spontaneous breathing occurs. Being a portable device, the neopuff can be susceptible to impact damage, for instance when accidentally dropped or subjected to considerable external force. It should be noted that the complaint device is more than 13 years old. Each neopuff unit is assembled and 100% tested in the production line to verify that each unit conforms to critical product specifications. Any unit that fails is rejected. The neopuff technical manual warns against dropping the neopuff or subjecting it to impact damage which may cause the unit to operate incorrectly. If the neopuff is suspected to have been damaged, the manometer and valve system should be performance tested. The neopuff technical manual states the following: dropping the neopuff / perivent infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit. If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient.
Patient Sequence No: 1, Text Type: N, H10


[187414639] A healthcare facility in (b)(6) reported that the gas inlet and outlet port of a rd900 neopuff infant resuscitator was broken. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611451-2020-00297
MDR Report Key9863022
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-03-20
Date of Report2020-02-20
Date Mfgr Received2020-02-20
Device Manufacturer Date2006-08-26
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS FARANAK GOMAROONI
Manufacturer Street173 TECHNOLOGY DRIVE SUITE 100
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9194534000
Manufacturer G1FISHER & PAYKEL HEALTHCARE LTD
Manufacturer Street15 MAURICE PAYKEL PLACE EAST TAMAKI
Manufacturer CityAUCKLAND, 2013
Manufacturer CountryNZ
Manufacturer Postal Code2013
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEOPUFF INFANT RESUSCITATOR
Generic NameBTL
Product CodeBTL
Date Received2020-03-20
Model NumberRD900
Catalog NumberRD900
Lot Number060826
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFISHER & PAYKEL HEALTHCARE LTD
Manufacturer Address15 MAURICE PAYKEL PLACE EAST TAMAKI AUCKLAND, 2013 NZ 2013


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.