[184482276] Product analysis :analysis was able to confirm the customer comment that the upper display hinges were broken. Analysis also identified excessive corrosion /contamination throughout the main case. Device was scrapped due to extensive corrosion throughout the main case. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[184482277] It was reported that mechanism that holds the top screen of the programmer was broken. The status of the device is expected to be re turned. There was no patient involvement. It was further reported that the screen hangs down and is not adjustable. The programmer was returned for analysis.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182208-2020-00554
MDR Report Key9864495
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-23
Date of Report2020-03-23
Date of Event2020-02-05
Date Mfgr Received2020-02-26
Device Manufacturer Date2002-07-25
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC, INC.
Manufacturer Street8200 CORAL SEA STREET NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal Code55112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK PROGRAMMER
Generic NamePROGRAMMER, PACEMAKER
Product CodeKRG
Date Received2020-03-23
Returned To Mfg2020-02-10
Model Number2090W
Catalog Number2090W
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC.
Manufacturer Address8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
12067 2020-03-23