MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,s report with the FDA on 2020-03-23 for MICRA MC1VR01 manufactured by Medtronic Ireland.
[184430576]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[184430577]
It was reported that the high thresholds and intermittent capture were observed one month post implant procedure of the leadless implantable pulse generator (ipg). Micro-dislodgment of the leadless ipg was suspected. The device was reprogrammed, however surgical revision of the device is planned. Leadless ipg remains in use. The patient is a participant in the post approval clinical surveillance product surveillance registry. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 9612164-2020-01289 |
MDR Report Key | 9864659 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,S |
Date Received | 2020-03-23 |
Date of Report | 2020-03-23 |
Date of Event | 2020-03-05 |
Date Mfgr Received | 2020-03-14 |
Date Added to Maude | 2020-03-23 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAULA BIXBY |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635055378 |
Manufacturer G1 | MEDTRONIC IRELAND |
Manufacturer Street | PARKMORE BUSINESS PARK WEST |
Manufacturer City | GALWAY |
Manufacturer Country | IE |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | MICRA |
Generic Name | LEADLESS PACEMAKER |
Product Code | PNJ |
Date Received | 2020-03-23 |
Model Number | MC1VR01 |
Catalog Number | MC1VR01 |
Operator | LAY USER/PATIENT |
Device Availability | N |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC IRELAND |
Manufacturer Address | PARKMORE BUSINESS PARK WEST GALWAY IE |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-03-23 |