MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-23 for GRASPIT M0063204010 320-401 manufactured by Boston Scientific Corporation.
[184457600]
The complainant was unable to provide the device lot number; therefore, the lot expiration and device manufacture dates are unknown. (b)(4). According to the complainant, the suspect device has been disposed and is not available for return. If any further relevant information is received, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10
[184457601]
It was reported to boston scientific corporation that a graspit nitinol stone retrieval forceps was used in the ureter during a lithotripsy procedure for removal of renal stones performed on (b)(6) 2020. According to the complainant, when trying to remove a lasered stone during the procedure the forceps broke at the tip were the forceps open. The fractured forceps was removed from the patient with another grasping forceps. However, the ureter was punctured from multiple attempts to remove the broken forceps. An attempt to insert a nephrostomy failed, and bilateral retroperitoneal drains were inserted instead. Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005099803-2020-01144 |
MDR Report Key | 9865234 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-23 |
Date of Report | 2020-03-23 |
Date of Event | 2020-02-25 |
Date Mfgr Received | 2020-02-25 |
Date Added to Maude | 2020-03-23 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | CAROLE MORLEY |
Manufacturer Street | 300 BOSTON SCIENTIFIC WAY |
Manufacturer City | MARLBOROUGH MA 01752 |
Manufacturer Country | US |
Manufacturer Postal | 01752 |
Manufacturer Phone | 5086834015 |
Manufacturer G1 | BOSTON SCIENTIFIC CORPORATION |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | GRASPIT |
Generic Name | DISLODGER, STONE, BASKET, URETERAL, METAL |
Product Code | FFL |
Date Received | 2020-03-23 |
Model Number | M0063204010 |
Catalog Number | 320-401 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BOSTON SCIENTIFIC CORPORATION |
Manufacturer Address | 300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-03-23 |