[188029334] It was report that patient has a full revision due to low impedance seen during revision surgery. The generator was replaced prophylactically while the lead was replaced due to low impedance. The low impedance was attributed to lead damage since visible damage was seen when lead was explanted. Dhr was reviewed and the device passed all functional specifications and quality tests and were sterilized prior to distribution. Explanted product has not been received to date. No additional relevant information has been received to date.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1644487-2020-00467
MDR Report Key9865267
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-23
Date of Report2020-03-23
Date of Event2020-02-27
Date Mfgr Received2020-02-27
Device Manufacturer Date2007-07-16
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. RACHEL KOHN
Manufacturer Street100 CYBERONICS BLVD SUITE 600
Manufacturer CityHOUSTON TX 77058
Manufacturer CountryUS
Manufacturer Postal77058
Manufacturer Phone2812287200
Manufacturer G1CYBERONICS - HOUSTON
Manufacturer Street100 CYBERONICS BLVD SUITE 600
Manufacturer CityHOUSTON TX 77058
Manufacturer CountryUS
Manufacturer Postal Code77058
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEAD MODEL 302
Generic NameLEAD
Product CodeLYJ
Date Received2020-03-23
Model Number302-20
Lot Number200551
Device Expiration Date2010-06-07
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCYBERONICS - HOUSTON
Manufacturer Address100 CYBERONICS BLVD HOUSTON TX 77058 US 77058


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-23

© 2026 FDA.report