O1 9733346

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-23 for O1 9733346 manufactured by Medtronic Navigation, Inc (littleton).

Event Text Entries

[184480585] The manufacturer representative went to the site to test the imaging system. The reported issue was confirmed and the bulkhead on the back of the mobile view station (mvs) was replaced and secured with the additional cable provided with the mvs ethernet cable. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[184480586] Medtronic received information regarding an imaging system being used in a sacroiliac and thoracolumbar procedure. It was reported that the communication between the imaging system and the navigation system was not solid. It was only solid when the ethernet cable was pushed and maintained straight on the back of the mobile view station (mvs). There was less than an hour delay in the case. No impact on patient outcome.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004785967-2020-00388
MDR Report Key9865332
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-23
Date of Report2020-03-23
Date of Event2020-03-10
Date Mfgr Received2020-03-10
Device Manufacturer Date2009-04-07
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Street300 FOSTER ST
Manufacturer CityLITTLETON MA 01460
Manufacturer CountryUS
Manufacturer Postal Code01460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameO1
Generic NameIMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Product CodeOXO
Date Received2020-03-23
Model Number9733346
Catalog Number9733346
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Address300 FOSTER ST LITTLETON MA 01460 US 01460


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-23

© 2026 FDA.report