CARELINK ENCORE PROGRAMMER 29901A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-23 for CARELINK ENCORE PROGRAMMER 29901A manufactured by Plexus Manufacturing Sdn. Bhd.

Event Text Entries

[184469665] Product analysis: analysis was able to confirm the customer comment that the programmer would not boot up past the logo screen until the screen was touched and that the cursor does not move. It was noted that the programmer was able to interrogate but not able to select however it did work with a known good display. Replaced computer platform (cp), reloaded and updated software and the device then passed all final functional and systems tests. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[184469666] It was reported that the programmer required multiple reboots before it could be used. It booted up successfully after three reboots but would then only interrogate the device. The screen was unresponsive to the stylus so the user was unable to program the device. The user had to reboot the programmer again. The programmer was returned for service. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004593495-2020-00304
MDR Report Key9865522
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-23
Date of Report2020-03-23
Date of Event2020-03-03
Date Mfgr Received2020-03-03
Device Manufacturer Date2017-12-20
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK ENCORE PROGRAMMER
Generic NamePROGRAMMER, PACEMAKER
Product CodeKRG
Date Received2020-03-23
Returned To Mfg2020-03-06
Model Number29901A
Catalog Number29901A
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-23

© 2024 FDA.report