INTELLANAV OPEN-IRRIGATED 87045

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-03-23 for INTELLANAV OPEN-IRRIGATED 87045 manufactured by Boston Scientific Corporation.

Event Text Entries

[184539406] It was reported that the catheter was sensing temperature lower than expected. During an ablation procedure for typical flutter, the physician inserted the intellanav open-irrigated catheter into the chamber. Distal temperature was about 31-32 degree celsius without ablation. With ablation energy on at 30 watts in power control mode, the temperature was about 19-20 degree celsius. The pump flow setting was 25-30 ml/min. The temperature sensor was under detecting. No error codes were displayed. No patient complications occurred. The procedure was completed with a new catheter.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-03692
MDR Report Key9865995
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-03-23
Date of Report2020-03-23
Date of Event2020-02-21
Date Mfgr Received2020-02-24
Device Manufacturer Date2019-06-24
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTIMOTHY DEGROOT
Manufacturer Street4100 HAMLINE AVENUE NORTH DC A330
Manufacturer CitySAINT PAUL MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone6515826168
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street302 PARKWAY, GLOBAL PARK
Manufacturer CityLA AURORA - HEREDIA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINTELLANAV OPEN-IRRIGATED
Generic NameINTELLANAV OI
Product CodeOAD
Date Received2020-03-23
Returned To Mfg2020-03-17
Model Number87045
Catalog Number87045
Lot Number0023992720
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-23