CTS22, 12X100 KII SLEEVE ZTHR 12/BX 101244801

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-23 for CTS22, 12X100 KII SLEEVE ZTHR 12/BX 101244801 manufactured by Applied Medical Resources.

Event Text Entries

[184498668] The event device is anticipated to be returned to applied medical for evaluation. A follow-up report will be provided upon completion of investigation.
Patient Sequence No: 1, Text Type: N, H10


[184498669] Procedure performed: laparoscopic pleurodesis. Event description: rep was not present for the case. The surgeon is a newly practicing surgeon. The surgeon placed the trocar between the rib cage and diaphragm. While looking around with the scope he noticed a broken fragment from the tip of the cannula. The case was converted to a open procedure to retrieve the plastic piece in the patient. The patient status is fine. The trocar is available to be returned. Intervention: converted to open to retrieve the plastic piece of the trocar. Patient status: patient is fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027111-2020-00433
MDR Report Key9866461
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-23
Date of Report2020-03-23
Date of Event2020-03-18
Date Mfgr Received2020-03-18
Device Manufacturer Date2019-11-14
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactWENDY KOBAYASHI
Manufacturer Street22872 AVENIDA EMPRESA
Manufacturer CityRANCHO SANTA MARGARITA CA 92688
Manufacturer CountryUS
Manufacturer Postal92688
Manufacturer Phone9497138059
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCTS22, 12X100 KII SLEEVE ZTHR 12/BX
Generic NameLAPAROSCOPE, GENERAL & PLASTIC SURGERY
Product CodeGCJ
Date Received2020-03-23
Model NumberCTS22
Catalog Number101244801
Lot Number1374795
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerAPPLIED MEDICAL RESOURCES
Manufacturer Address22872 AVENIDA EMPRESA RANCHO SANTA MARGARITA CA 92688 US 92688


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-23

© 2026 FDA.report