MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-23 for SJM TRIFECTA VALVE TF-21A manufactured by St. Jude Medical, Costa Rica Ltda (cvd).
[184710217]
The results/method and conclusion codes along with investigation results will be provided in the final report.
Patient Sequence No: 1, Text Type: N, H10
[184710218]
Reference manufacturing report number 3007113487-2020-00005. On an unknown date, an abbott mechanical aortic valve was implanted. On (b)(6) 2015, mechanical valve was explanted due to stenosis that led to obstruction of flow and was replaced with a 21mm trifecta valve. On an unknown date, stenosis and perivalvular leak was observed on the trifecta valve. On (b)(6) 2020, the valve was explanted and replaced with a 21mm biological edwards valve. Pannus and calcification was observed on the explanted trifecta valve. Yh (b)(6) 2020.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3008452825-2020-00162 |
MDR Report Key | 9866495 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-23 |
Date of Report | 2020-03-23 |
Date of Event | 2020-02-24 |
Date Mfgr Received | 2020-02-24 |
Device Manufacturer Date | 2014-10-07 |
Date Added to Maude | 2020-03-23 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAMELA YIP |
Manufacturer Street | 5050 NATHAN LANE N |
Manufacturer City | PLYMOUTH MN 55442 |
Manufacturer Country | US |
Manufacturer Postal | 55442 |
Manufacturer Phone | 6517565400 |
Manufacturer G1 | ST. JUDE MEDICAL, COSTA RICA LTDA (CVD) |
Manufacturer Street | PARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL |
Manufacturer City | ALAJUELA, COSTA RICA 01897-405 |
Manufacturer Country | CS |
Manufacturer Postal Code | 01897-4050 |
Single Use | 3 |
Remedial Action | RL |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SJM TRIFECTA VALVE |
Generic Name | HEART-VALVE, NON-ALLOGRAFT TISSUE |
Product Code | LWR |
Date Received | 2020-03-23 |
Model Number | TF-21A |
Catalog Number | TF-21A |
Lot Number | 4847765 |
Device Expiration Date | 2016-10-05 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ST. JUDE MEDICAL, COSTA RICA LTDA (CVD) |
Manufacturer Address | PARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL ALAJUELA, COSTA RICA 01897-405 CS 01897-4050 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other; 2. Required No Informationntervention | 2020-03-23 |