C-STEM SZ2 9611-62-000 961162000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-03-23 for C-STEM SZ2 9611-62-000 961162000 manufactured by Depuy International Ltd - 8010379.

Event Text Entries

[188860732] (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[188860733] The primary surgery was performed on (b)(6) 2004 via tha. It was reported that the revision surgery was performed on (b)(6) 2020 by replacing the c-stem (p/n: unknown) and the cup (p/n: unknown) due to a pain which the patient had left and right side of his hip joints at the same time in autumn 2019 (the date was unknown). The revision surgery was completed successfully, and it was unknown whether there was surgical delay or not. The condition after surgery was good. The surgeon found that there was something like a corrosion from the middle point to distal part of the explanted stem. He thought that this caused the pain. No further information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1818910-2020-08910
MDR Report Key9868038
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-03-23
Date of Report2020-03-05
Date of Event2020-02-10
Date Mfgr Received2020-03-05
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street700 ORTHOPAEDIC DR.
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal465810988
Manufacturer Phone6107428552
Manufacturer G1LEEDS MFG & MATERIAL WAREHOUSE
Manufacturer StreetST ANTHONYS ROAD
Manufacturer CityLEEDS LS118DT
Manufacturer CountryUK
Manufacturer Postal CodeLS11 8DT
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameC-STEM SZ2
Generic NameC-STEM HIP IMPLANT : HIP FEMORAL STEM
Product CodeKXA
Date Received2020-03-23
Model Number9611-62-000
Catalog Number961162000
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY INTERNATIONAL LTD - 8010379
Manufacturer AddressST. ANTHONY'S ROAD LEEDS LS118DT UK LS11 8DT


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-23

© 2026 FDA.report