SYNCHRO2-14 SOFT STRAIGHT 200CM M00326010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-23 for SYNCHRO2-14 SOFT STRAIGHT 200CM M00326010 manufactured by Stryker Neurovascular-utah-salt Lake City.

MAUDE Entry Details

Report Number3012931345-2020-00045
MDR Report Key9868271
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-23
Date of Report2020-03-23
Date of Event2020-03-03
Date Mfgr Received2020-03-03
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TARA LOPEZ
Manufacturer Street47900 BAYSIDE PARKWAY
Manufacturer CityFREMONT CA 94538
Manufacturer CountryUS
Manufacturer Postal94538
Manufacturer Phone5104132500
Manufacturer G1STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
Manufacturer Street4870 WEST 2100 SOUTH
Manufacturer CitySALT LAKE CITY UT 84120
Manufacturer CountryUS
Manufacturer Postal Code84120
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYNCHRO2-14 SOFT STRAIGHT 200CM
Generic NameWIRE, GUIDE, CATHETER
Product CodeDQX
Date Received2020-03-23
Returned To Mfg2020-03-13
Model NumberM00326010
Catalog NumberM00326010
Lot Number0000039309
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
Manufacturer Address4870 WEST 2100 SOUTH SALT LAKE CITY UT 84120 US 84120


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.