PROSA VALVE FV701T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-23 for PROSA VALVE FV701T manufactured by Christoph Miethke Gmbh & Co. Kg.

Event Text Entries

[188553529] Height: 180cm investigation: visual inspection: a deformation of the outer housing of the prosa valve was observed through the visual inspection. The prosa valve housing was subsequently measured and confirmed the presence of a deformation. The housing deformation measured at -0. 0895 mm, outside the tolerance of 0? 0. 02 mm. Permeability test: a permeability test has shown that the prosa valve is permeable. Adjustment test: the prosa valve was tested and could be adjusted in all pressure ranges without any form of force being required. The brake is damaged is adjustable without to all specified pressures. Braking force and brake function test: the brake functionality test has shown that the brake function did not operate as expected. The results indicate that the rotor no longer performs as required. Results: first, we performed a visual inspection of the prosa valve. A deformation of the outer housing of the prosa valve was observed through the visual inspection. This deformation was confirmed through a measurement of the plan parallelity. Next we tested the permeability, adjustability, the brake functionality and brake force. The valve is permeable, but the brake is damaged and the braking force is no longer available. Finally, we have dismantled the valve. There were no visible deposits observed inside the valve. Based on our investigation results, we can confirm the suspicion that "the valve adjusts by itself" at the time of our investigation. We assume that the observed deformation is the cause of the functional impairment of the adjustability. The cause of the deformation of the prosa valve and the resultant defect of the rotor could not be determined through our investigation. Significant outside pressure, for example by too much force from the prosa adjustment tool or by a fall or impact to the head of the patient, can compromise the integrity of the valve. We can exclude a defect at the time of release. The valve met all specifications of the final inspection when released from christoph miethke (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[188553530] It was reported that there is an issue with valve. The reporter indicated that a post-operative valve does not hold pressure. The device was explanted. Additional event details is not available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004721439-2020-00068
MDR Report Key9868384
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-23
Date of Report2020-03-23
Date of Event2020-03-06
Date Mfgr Received2020-03-09
Device Manufacturer Date2019-11-01
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJOERG KNEBEL
Manufacturer Street2 ULANENWEG
Manufacturer CityPOTSDAM D, 14469
Manufacturer CountryGM
Manufacturer Postal14469
Manufacturer G1CHRISTOPH MIETHKE GMBH & CO. KG
Manufacturer Street2 ULANENWEG
Manufacturer CityPOTSDAM D, 14469
Manufacturer CountryGM
Manufacturer Postal Code14469
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROSA VALVE
Generic NameHYDROCEPHALUS MANAGEMENT
Product CodeJXG
Date Received2020-03-23
Returned To Mfg2020-03-09
Model NumberFV701T
Catalog NumberFV701T
Lot Number20044492
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHRISTOPH MIETHKE GMBH & CO. KG
Manufacturer Address2 ULANENWEG POTSDAM D, 14469 GM 14469


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.