MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-23 for DUAL CHAMBER TEMPORARY PACEMAKER 5392 manufactured by Plexus Manufacturing Sdn. Bhd.
[184638395]
Product analysis: analysis was able to confirm the customer comments that the external pulse generator (epg) had an issue with the ventricular output connector, was emitting a rattling noise and had knobs missing. It was noted that the ventricular output connector was broken, two upper case bosses were broken resulting in debris making rattling noise, and two encoder knobs were missing. It was further noted that the hanger assembly was cracked, lower case was broken, main label and manufacturing label were damaged and insulator label was contaminated. All found defective parts were replaced and all other identified issues were resolved with firmware up dated and the epg passed all final functional tests. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[184638396]
It was reported that the external pulse generator (epg) had an issue with the ventricular output connector, was emitting a rattling noise and had knobs missing. The epg was received into service and repair. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3004593495-2020-00308 |
MDR Report Key | 9869141 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2020-03-23 |
Date of Report | 2020-03-23 |
Date of Event | 2020-02-24 |
Date Mfgr Received | 2020-02-24 |
Device Manufacturer Date | 2017-07-28 |
Date Added to Maude | 2020-03-23 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAULA BIXBY |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635055378 |
Manufacturer G1 | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Street | BAYAN LEPAS FREE INDUSTRIAL ZO |
Manufacturer City | BAYAN LEPAS 11900 |
Manufacturer Country | MY |
Manufacturer Postal Code | 11900 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | DUAL CHAMBER TEMPORARY PACEMAKER |
Generic Name | PULSE-GENERATOR, PACEMAKER, EXTERNAL |
Product Code | DTE |
Date Received | 2020-03-23 |
Returned To Mfg | 2020-02-27 |
Model Number | 5392 |
Catalog Number | 5392 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Address | BAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-23 |