AMPLATZER SEPTAL OCCLUDER 9-ASD-008

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-23 for AMPLATZER SEPTAL OCCLUDER 9-ASD-008 manufactured by Aga Medical Corporation.

Event Text Entries

[184662412] Further information regarding this event has been requested. The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
Patient Sequence No: 1, Text Type: N, H10


[184662413] On (b)(6) 2020, a 8mm amplatzer septal occluder (9-asd-008 lot #7017005) was selected for implant to close an asd in the patient. Upon deployment the left atrial disk was "cobra" maligned. The physician tried several different techniques to get the device to deploy but was unsuccessful. The device was resheathed and removed from the patient. The physician exchanged the device for a 8mm amplatzer septal occluder (9-asd-008 lot #7054723) and successfully closed the defect with no problem. There was no clinically significant delay in the procedure and the patient remained hemodynamically stable throughout the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2135147-2020-00097
MDR Report Key9869456
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-23
Date of Report2020-03-30
Date of Event2020-02-24
Date Mfgr Received2020-03-24
Device Manufacturer Date2019-05-28
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAMELA YIP
Manufacturer Street5050 NATHAN LANE N
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1AGA MEDICAL CORPORATION
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal Code55442
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMPLATZER SEPTAL OCCLUDER
Generic NameTRANSCATHETER SEPTAL OCCLUDER
Product CodeMLV
Date Received2020-03-23
Returned To Mfg2020-03-16
Model Number9-ASD-008
Catalog Number9-ASD-008
Lot Number7017005
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAGA MEDICAL CORPORATION
Manufacturer Address5050 NATHAN LANE NORTH PLYMOUTH MN 55442 US 55442


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.