ENROUTE NEUROPROTECTION SYSTEM SR-200-NPS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-23 for ENROUTE NEUROPROTECTION SYSTEM SR-200-NPS manufactured by Silk Road Medical.

Event Text Entries

[189082438] The product associated with this complaint was not returned to the manufacturer for analysis. A review of manufacturer records for this device was completed and no issues were identified that could have lead to the adverse event reported. Complaints will continue to be reviewed and monitored for trends.
Patient Sequence No: 1, Text Type: N, H10


[189082439] A patient underwent a transcarotid arterial revascularization procedure. The physician used the "stop short" technique when introducing the arterial sheath due to the calcium in the carotid bifurcation. Upon introduction of the guidewire, a dissection occurred. The decision was made to convert the procedure to a carotid endarterectomy, during which time the dissection was surgically repaired. There were no complications and the patient was stable at the end of the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3014526664-2020-00049
MDR Report Key9869634
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-23
Date of Report2020-03-23
Date of Event2020-03-03
Date Mfgr Received2020-03-03
Device Manufacturer Date2019-07-22
Date Added to Maude2020-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SHARON RIDEOUT
Manufacturer Street1213 INNSBRUCK DRIVE
Manufacturer CitySUNNYVALE, CA
Manufacturer CountryUS
Manufacturer Phone7209002
Manufacturer G1SILK ROAD MEDICAL
Manufacturer Street1213 INNSBRUCK DRIVE
Manufacturer CitySUNNYVALE, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENROUTE NEUROPROTECTION SYSTEM
Generic NameNPS
Product CodeNTE
Date Received2020-03-23
Model NumberSR-200-NPS
Catalog NumberSR-200-NPS
Lot Number300898
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSILK ROAD MEDICAL
Manufacturer Address1213 INNSBRUCK DRIVE SUNNYVALE, CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-03-23

© 2026 FDA.report