BARD-PARKER? SAFESWITCH? REUSABLE SCALPEL HANDLE, SIZE 7, NON-STERILE ST-1017NS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-24 for BARD-PARKER? SAFESWITCH? REUSABLE SCALPEL HANDLE, SIZE 7, NON-STERILE ST-1017NS manufactured by Aspen Surgical Products, Caledonia.

Event Text Entries

[186724555] Aspen surgical received a report from the end user indicating that during a procedure, the scalpel came off of the scalpel handle which required retrieval. No sample or photographic was provided. No adverse effect to the patient was reported. A manufacturing lot number was provided for review. As reported by the end user, during a shoulder procedure the blade came off the handle inside the patient. The blade required retrieval. The doctor had inserted the blade/handle into the shoulder to make an incision and when pulling the handle out of the body, the blade came off the handle inside the patient. Patient was not harmed. A review of the device history record showed no non-conformance's related to the reported issue. Actual device in use was not returned, however remaining units from package were tested with no blade retention issues. Based on remaining units and lot number review, cause is not establish. No further information is available on the product at this time. The device in use and photographic evidence was not available. However if any additional relevant information is identified or sample is returned, the additional relevant information will be submitted in a supplemental report. Device not returned.
Patient Sequence No: 1, Text Type: N, H10


[186724556] Aspen surgical received a report from the end user indicating that during a procedure, the scalpel came off of the scalpel handle which required retrieval. No adverse effect to the patient was reported. The device was in use during the alleged malfunction/event. This report was filed in our complaint handling system as number c(b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1836161-2020-00016
MDR Report Key9871502
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-03-24
Date of Report2020-02-27
Date of Event2020-02-26
Date Mfgr Received2020-02-27
Date Added to Maude2020-03-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JORDAN HACKERT
Manufacturer Street6945 SOUTHBELT DR. S.E.
Manufacturer CityCALEDONIA MI 49316
Manufacturer CountryUS
Manufacturer Postal49316
Manufacturer Phone6165367508
Manufacturer G1SAME AS ABOVE
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBARD-PARKER? SAFESWITCH? REUSABLE SCALPEL HANDLE, SIZE 7, NON-STERILE
Generic NameSCALPEL HANDLE
Product CodeGDZ
Date Received2020-03-24
Model NumberST-1017NS
Lot Number146965
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerASPEN SURGICAL PRODUCTS, CALEDONIA
Manufacturer Address6945 SOUTHBELT DR. S.E. CALEDONIA MI 49316 US 49316


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-24