ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM 24653

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-03-24 for ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM 24653 manufactured by Boston Scientific Corporation.

Event Text Entries

[184805873] Age at time of event: 18 years or older.
Patient Sequence No: 1, Text Type: N, H10


[184805954] It was reported that partial stent deployment occurred. A 6x120x130 eluvia self-expanding stent was selected for use in the 100% stenosed, non tortuous, and severe calcified superficial femoral artery (sfa). A contralateral approach was used and the lesion was pre-dilated with a 4mm sterling balloon. The angle of aorta was tight. The stent was advanced over a 0. 014 jupiter guidewire without replacing with 0. 035 guidewire inside a non-bsc 6 fr sheath. Stent deployment was initiated. When deployed about 6 cm, the thumbwheel stopped rotating, and the stent remained on the shaft. Then, the physician broke the handle, and pulled the inner shaft inside, and the stent was placed. There was section at about 1cm in the middle of the stent that the struts were wide open and the struts were stretched. There were no patient complications and the patient's condition was good post procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-03679
MDR Report Key9871656
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-03-24
Date of Report2020-03-24
Date of Event2020-03-04
Date Mfgr Received2020-03-04
Date Added to Maude2020-03-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC IRELAND LIMITED
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Generic NameSTENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Product CodeNIU
Date Received2020-03-24
Model Number24653
Catalog Number24653
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-24