WALLSTENT ENDOPROSTHESIS 26280

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-24 for WALLSTENT ENDOPROSTHESIS 26280 manufactured by Boston Scientific Corporation.

Event Text Entries

[184674751] It was reported that stent migration occurred and required additional intervention. On (b)(6) 2020, the more than 75% stenosed target lesion, measuring 15mm in healthy vessel, was located in the right subclavian vein and noted via ivus. During the procedure, a 16mm x 60mm x 75cm wallstent endoprosthesis stent was deployed. The stent was pre and post dilated with charger balloon dilatation catheters 8mm x 40mm and 10 mm x 40mm. The physician was pleased with the results and the patient outcome was optimal. On (b)(6) 2020, it was noted that the wallstent had migrated to the superior vena cava, past the initial stenosis and placement by at least 2-3 cm. A 16mm x 60mm vici venous stent was placed at the lesion. Post dilatation was completed with a balloon catheter. There were no patient complications nor injuries reported and the patient was stable and doing well post procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-03743
MDR Report Key9871793
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-24
Date of Report2020-03-24
Date of Event2020-03-12
Date Mfgr Received2020-03-12
Device Manufacturer Date2019-03-27
Date Added to Maude2020-03-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWALLSTENT ENDOPROSTHESIS
Generic NameCATHETER, BILIARY, DIAGNOSTIC
Product CodeMAF
Date Received2020-03-24
Model Number26280
Catalog Number26280
Lot Number0023550585
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
1121. Required No Informationntervention 2020-03-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.