[184670295] Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[184670296] A surgeon reported a patient experienced a corneal burn during a cataract extraction procedure. The patient had a small anterior chamber, therefore it was necessary to use a greater amount of ophthalmic viscoelastic (ovd) to inflate the anterior chamber. The phacoemulsification (phaco) tip filled with the ovd upon insertion of the tip into the eye. The ovd was crystallized upon activation of the handpiece and the phaco tip clogged. Without cooling the burn was immediate. The occlusion alarm from the system did sound and was heard by the surgeon, but he did not realize the severity until the burn was noted. The tip was removed form the eye, rinsed and unclogged and the case was completed. A suture was required to close the wound. On post operative day one the patient presented with endothelial detachment and corneal folds. Additional information has been requested but not received.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523835-2020-00072
MDR Report Key9872067
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-24
Date of Report2020-03-24
Date of Event2020-02-01
Date Mfgr Received2020-02-27
Date Added to Maude2020-03-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN SCHLECH
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175514979
Manufacturer G1ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
Manufacturer Street714 COLUMBIA AVENUE
Manufacturer CitySINKING SPRING PA 19608
Manufacturer CountryUS
Manufacturer Postal Code19608
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NamePHACO TIP
Generic NameUNIT, PHACOFRAGMENTATION
Product CodeHQC
Date Received2020-03-24
Model NumberNA
Catalog NumberASKU
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LLC - ALCON PRECISION DEVICE
Manufacturer Address714 COLUMBIA AVENUE SINKING SPRING PA 19608 US 19608


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.