TRITANIUM BPLATE TRIATHLON S4 5536-B-400 5536B400

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other report with the FDA on 2020-03-24 for TRITANIUM BPLATE TRIATHLON S4 5536-B-400 5536B400 manufactured by Stryker Orthopaedics-mahwah.

Event Text Entries

[185044063] It was noted that the device is not available for evaluation. Should additional information become available, it will be provided in a supplemental report upon completion of the investigation. Device not returned.
Patient Sequence No: 1, Text Type: N, H10


[185044066] Patient reporting painful tka. Suspected loose tritanium tibial component. Patient was revised.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002249697-2020-00588
MDR Report Key9872143
Report SourceHEALTH PROFESSIONAL,OTHER
Date Received2020-03-24
Date of Report2020-03-24
Date of Event2020-02-25
Date Mfgr Received2020-02-25
Date Added to Maude2020-03-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. COLLIN NEITZEL
Manufacturer Street325 CORPORATE DRIVE
Manufacturer CityMAHWAH NJ 07430
Manufacturer CountryUS
Manufacturer Postal07430
Manufacturer Phone2018315000
Manufacturer G1STRYKER ORTHOPAEDICS-LIMERICK
Manufacturer StreetRAHEEN BUSINESS PARK
Manufacturer CityLIMERICK NA
Manufacturer CountryIE
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRITANIUM BPLATE TRIATHLON S4
Generic NamePROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Product CodeMBH
Date Received2020-03-24
Model Number5536-B-400
Catalog Number5536B400
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ORTHOPAEDICS-MAHWAH
Manufacturer Address325 CORPORATE DRIVE MAHWAH NJ 07430 US 07430


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-03-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.