TECNIS 1-PIECE ZCB00 ZCB0000190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-24 for TECNIS 1-PIECE ZCB00 ZCB0000190 manufactured by Johnson & Johnson Surgical Vision, Inc..

Event Text Entries

[184750198] Additional information: if implanted; give date: not applicable as the iol was not implanted. If explanted; give date: not applicable as the iol was not implanted. Device evaluation: the lens was received in the original daisy wheel. Additionally, the folding carton and complaint shipping guide were received. Visual inspection with the unaided eye revealed viscoelastic residue on the optic body and haptics, and that the lens was received cut in half, which is consistent with a lens that was handled during explant. Based on the return condition of the lens no product evaluation could be performed and the? Particle? Observed by the customer could not be confirmed. The complaint issue could not be confirmed, and no product deficiency was identified. Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification. Historical data analysis: a search revealed that no other complaints for this production order number have been received. Conclusion: as a result, of the investigation there is no indication of a product malfunction or product quality deficiency. The reported issue was not verified. All pertinent information available to johnson & johnson surgical vision, inc. Has been submitted.
Patient Sequence No: 1, Text Type: N, H10


[184750199] After zcb00 19. 0 diopter intraocular lens (iol) was inserted in the patient? S ocular sinister (left eye) customer account reported a particle was observed on the lens. It was also reported there was no patient injury and the procedure was completed using another zcb00 19. 0 diopter lens. Patient outcome was reported as fine. No further information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2648035-2020-00272
MDR Report Key9874461
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-03-24
Date of Report2020-03-24
Date of Event2020-02-24
Report Date2005-01-01
Date Reported to FDA2005-01-01
Date Reported to Mfgr2005-01-10
Date Mfgr Received2020-02-24
Date Added to Maude2020-03-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSOMYATA NAGPAL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478200
Manufacturer G1JOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer StreetROAD 402 NORTH, KM 4.2 ANASCO INDUSTRIAL PARK
Manufacturer CityANASCO PR 00610
Manufacturer CountryUS
Manufacturer Postal Code00610
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameTECNIS 1-PIECE
Generic NameMONOFOCAL IOLS
Product CodeHQL
Date Received2020-03-24
Returned To Mfg2020-03-12
Model NumberZCB00
Catalog NumberZCB0000190
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerJOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer Address1700 E. ST. ANDREW PLACE SANTA ANA CA 92705 US 92705

Device Sequence Number: 101

Product Code---
Date Received2020-03-24
Device Sequence No101
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-24