MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-24 for TECNIS 1-PIECE ZCB00 ZCB0000190 manufactured by Johnson & Johnson Surgical Vision, Inc..
[184750198]
Additional information: if implanted; give date: not applicable as the iol was not implanted. If explanted; give date: not applicable as the iol was not implanted. Device evaluation: the lens was received in the original daisy wheel. Additionally, the folding carton and complaint shipping guide were received. Visual inspection with the unaided eye revealed viscoelastic residue on the optic body and haptics, and that the lens was received cut in half, which is consistent with a lens that was handled during explant. Based on the return condition of the lens no product evaluation could be performed and the? Particle? Observed by the customer could not be confirmed. The complaint issue could not be confirmed, and no product deficiency was identified. Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification. Historical data analysis: a search revealed that no other complaints for this production order number have been received. Conclusion: as a result, of the investigation there is no indication of a product malfunction or product quality deficiency. The reported issue was not verified. All pertinent information available to johnson & johnson surgical vision, inc. Has been submitted.
Patient Sequence No: 1, Text Type: N, H10
[184750199]
After zcb00 19. 0 diopter intraocular lens (iol) was inserted in the patient? S ocular sinister (left eye) customer account reported a particle was observed on the lens. It was also reported there was no patient injury and the procedure was completed using another zcb00 19. 0 diopter lens. Patient outcome was reported as fine. No further information is available.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2648035-2020-00272 |
MDR Report Key | 9874461 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2020-03-24 |
Date of Report | 2020-03-24 |
Date of Event | 2020-02-24 |
Report Date | 2005-01-01 |
Date Reported to FDA | 2005-01-01 |
Date Reported to Mfgr | 2005-01-10 |
Date Mfgr Received | 2020-02-24 |
Date Added to Maude | 2020-03-24 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | SOMYATA NAGPAL |
Manufacturer Street | 1700 EAST ST. ANDREW PLACE |
Manufacturer City | SANTA ANA CA 92705 |
Manufacturer Country | US |
Manufacturer Postal | 92705 |
Manufacturer Phone | 7142478200 |
Manufacturer G1 | JOHNSON & JOHNSON SURGICAL VISION, INC. |
Manufacturer Street | ROAD 402 NORTH, KM 4.2 ANASCO INDUSTRIAL PARK |
Manufacturer City | ANASCO PR 00610 |
Manufacturer Country | US |
Manufacturer Postal Code | 00610 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TECNIS 1-PIECE |
Generic Name | MONOFOCAL IOLS |
Product Code | HQL |
Date Received | 2020-03-24 |
Returned To Mfg | 2020-03-12 |
Model Number | ZCB00 |
Catalog Number | ZCB0000190 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | JOHNSON & JOHNSON SURGICAL VISION, INC. |
Manufacturer Address | 1700 E. ST. ANDREW PLACE SANTA ANA CA 92705 US 92705 |
Product Code | --- |
Date Received | 2020-03-24 |
Device Sequence No | 101 |
Device Event Key | 0 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-24 |