LEVEL 1 TRAUMA FAST FLOW SYSTEM H-1200 8002950

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-24 for LEVEL 1 TRAUMA FAST FLOW SYSTEM H-1200 8002950 manufactured by Smiths Medical Asd, Inc.

Event Text Entries

[184815155] Evaluation results: one level 1 trauma fast flow system was returned for investigation in used condition. There was a crack in the return tube which had leaked water and damaged multiple components. Heater 2 insulation was also torn. The customer reported product problem was confirmed during testing. The temperature fluctuated. The return tube and main pcb were replaced as a result. The product problem was attributed to a design issue.
Patient Sequence No: 1, Text Type: N, H10


[184815156] It was reported that the device exhibited intermittent heating. No patient injury or complications were reported in relation to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-02220
MDR Report Key9875515
Report SourceUSER FACILITY
Date Received2020-03-24
Date of Report2020-03-24
Date Mfgr Received2020-02-25
Device Manufacturer Date2015-05-26
Date Added to Maude2020-03-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer G1SMITHS MEDICAL ASD, INC
Manufacturer Street3350 GRANADA AVENUE NORTH SUITE 100
Manufacturer CityOAKDALE, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEVEL 1 TRAUMA FAST FLOW SYSTEM
Generic NameWARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Product CodeBSB
Date Received2020-03-24
Returned To Mfg2019-10-10
Model NumberH-1200
Catalog Number8002950
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.