ZIMMER SKIN GRAFT MESHER 00770100000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2020-03-25 for ZIMMER SKIN GRAFT MESHER 00770100000 manufactured by Zimmer Surgical, Inc..

Event Text Entries

[186734815] This event has been recorded by zimmer biomet under (b)(4). Once the investigation is completed, a follow-up/final report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[186734816] It has been reported that the mesher was working backwards. The handle would not scratch it properly, and the gears possibly were not meshing. Hospital tried using the mesher during a case that required skin grafting, however the handle will not engage properly, and the skin graft carrier will go backwards. The event occurred during surgery. This piece of equipment is not used very often, and it is believed this was the first time we? Ve used it after it came back from its previous repair. No harm, no delay, no additional graft required. The procedure was delayed significantly(approximately 45-min,) due to the fact that hospital had to find a solution to make the mesher work, as the skin graft area was a large one, and we needed the skin meshed( manually putting slits in the skin was not an option because of the large graft needed). No adverse event was reported as a result of this malfunction.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001526350-2020-00310
MDR Report Key9876722
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2020-03-25
Date of Report2020-03-24
Date of Event2020-02-26
Date Mfgr Received2020-03-04
Date Added to Maude2020-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CHRISTINA ARNT
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5745273773
Manufacturer G1ZIMMER SURGICAL, INC.
Manufacturer Street200 WEST OHIO AVENUE
Manufacturer CityDOVER OH 44622
Manufacturer CountryUS
Manufacturer Postal Code44622
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameZIMMER SKIN GRAFT MESHER
Generic NameEXPANDER, SURGICAL, SKIN GRAFT
Product CodeFZW
Date Received2020-03-25
Catalog Number00770100000
Lot NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerZIMMER SURGICAL, INC.
Manufacturer Address200 WEST OHIO AVENUE DOVER OH 44622 US 44622


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-25

© 2024 FDA.report