TORCON NB ADVANTAGE PEDIATRIC ANGIOGRAPHIC CATHETER G13077 HNB5.0-38-110-P-12S-PED

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-25 for TORCON NB ADVANTAGE PEDIATRIC ANGIOGRAPHIC CATHETER G13077 HNB5.0-38-110-P-12S-PED manufactured by Cook Inc.

Event Text Entries

[185315145] Pma/510(k) number = pre-amendment. This report includes information known at this time. A follow-up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[185315146] As reported, during a procedure described as "coronary vessels diagnosis", the hub separated from a torcon nb advantage pediatric angiographic catheter. The catheter was connected to a cook high pressure connecting tube for power injection at 600 psi. During injection, the hub separated from the catheter shaft. A section of the device did not remain inside the patient? S body. The patient did not require any additional procedures due to this occurrence. According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2020-00681
MDR Report Key9877938
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-25
Date of Report2020-03-25
Date of Event2020-03-13
Date Mfgr Received2020-03-16
Device Manufacturer Date2019-12-09
Date Added to Maude2020-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. JENNIFER CANADA
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTORCON NB ADVANTAGE PEDIATRIC ANGIOGRAPHIC CATHETER
Generic NameDQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Product CodeDQO
Date Received2020-03-25
Model NumberG13077
Catalog NumberHNB5.0-38-110-P-12S-PED
Lot Number10181616
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-25

© 2024 FDA.report