STERLING 24690

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-03-25 for STERLING 24690 manufactured by Boston Scientific Corporation.

Event Text Entries

[185034837] It was reported that balloon rupture occurred. The 99% stenosed target lesion was located in the moderately tortuous and severely calcified superficial femoral artery. The percutaneous transluminal angioplasty was performed by contralateral approach. After crossing the lesion with a guidewire, a 6. 0mmx100mmx150cm (4f) sterling balloon catheter was advanced for dilatation. During procedure, the balloon ruptured at unknown pressure. The procedure was completed with a different device. No patient complications nor injuries were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-03849
MDR Report Key9878178
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-03-25
Date of Report2020-03-25
Date of Event2020-03-18
Date Mfgr Received2020-03-19
Device Manufacturer Date2019-01-25
Date Added to Maude2020-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal Code55311
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTERLING
Generic NameCATHETER, PERCUTANEOUS
Product CodeLIT
Date Received2020-03-25
Model Number24690
Catalog Number24690
Lot Number0023243172
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-25

© 2026 FDA.report