MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-25 for INFUSOR 2C1063KP manufactured by Baxter Healthcare Corporation.
[184863149]
The device was discarded; therefore, a device analysis could not be completed. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10
[184863150]
It was reported that a large volume infusor overinfused during patient infusion. The reporter stated that after the expected therapy time of 24 hours at 10ml/hr, the pump ran dry within 12 hours. The device had been filled with cephazolin in d5%. There was no patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5
[188469473]
Additional information was added: the lot was manufactured from june 12, 2019 - june 13, 2019. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1416980-2020-01702 |
MDR Report Key | 9878335 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-25 |
Date of Report | 2020-03-25 |
Date of Event | 2020-03-02 |
Date Mfgr Received | 2020-03-02 |
Date Added to Maude | 2020-03-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | PHARMACIST |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Street | 25212 W. ILLINOIS ROUTE 120 |
Manufacturer City | ROUND LAKE IL 60073 |
Manufacturer Country | US |
Manufacturer Postal | 60073 |
Manufacturer Phone | 2242702068 |
Manufacturer G1 | BAXTER HEALTHCARE - IRVINE |
Manufacturer Street | 17511 ARMSTRONG AVENUE BUILDING 3 |
Manufacturer City | IRVINE CA 92614 |
Manufacturer Country | US |
Manufacturer Postal Code | 92614 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | INFUSOR |
Generic Name | PUMP, INFUSION, ELASTOMERIC |
Product Code | MEB |
Date Received | 2020-03-25 |
Model Number | NA |
Catalog Number | 2C1063KP |
Lot Number | 19F031 |
Operator | LAY USER/PATIENT |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BAXTER HEALTHCARE CORPORATION |
Manufacturer Address | DEERFIELD IL |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-25 |