INFUSOR 2C1063KP

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-25 for INFUSOR 2C1063KP manufactured by Baxter Healthcare Corporation.

Event Text Entries

[184863149] The device was discarded; therefore, a device analysis could not be completed. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[184863150] It was reported that a large volume infusor overinfused during patient infusion. The reporter stated that after the expected therapy time of 24 hours at 10ml/hr, the pump ran dry within 12 hours. The device had been filled with cephazolin in d5%. There was no patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[188469473] Additional information was added: the lot was manufactured from june 12, 2019 - june 13, 2019. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416980-2020-01702
MDR Report Key9878335
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-25
Date of Report2020-03-25
Date of Event2020-03-02
Date Mfgr Received2020-03-02
Date Added to Maude2020-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - IRVINE
Manufacturer Street17511 ARMSTRONG AVENUE BUILDING 3
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal Code92614
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINFUSOR
Generic NamePUMP, INFUSION, ELASTOMERIC
Product CodeMEB
Date Received2020-03-25
Model NumberNA
Catalog Number2C1063KP
Lot Number19F031
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE CORPORATION
Manufacturer AddressDEERFIELD IL


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-25

© 2026 FDA.report