MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-25 for POLYFLUX 140H NI 955608 manufactured by Baxter Healthcare - Hechingen Baden Dialyzer.
[185296132]
The device was received for evaluation. A leakage test on the dialysate side was performed and a crack in the welding zone was found. The failure could be confirmed. The review of the welding parameters of the product showed no conspicuousness, they were within the specification. The cause was undetermined. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10
[185296133]
It was reported that ten minutes after starting treatment with a polyflux 140h, an external dialysate leakage was observed between the header and the housing. There was no patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 9611369-2020-00026 |
MDR Report Key | 9878442 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-25 |
Date of Report | 2020-03-25 |
Date of Event | 2020-02-29 |
Date Mfgr Received | 2020-02-29 |
Date Added to Maude | 2020-03-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Street | 25212 W. ILLINOIS ROUTE 120 |
Manufacturer City | ROUND LAKE IL 60073 |
Manufacturer Country | US |
Manufacturer Postal | 60073 |
Manufacturer Phone | 2242702068 |
Manufacturer G1 | BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER |
Manufacturer Street | HOLGER - CRAFOORD - STRASSE 26 |
Manufacturer City | HECHINGEN D-72379 |
Manufacturer Country | GM |
Manufacturer Postal Code | D-72379 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | POLYFLUX 140H |
Generic Name | DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM |
Product Code | KDI |
Date Received | 2020-03-25 |
Returned To Mfg | 2020-03-20 |
Model Number | NI |
Catalog Number | 955608 |
Lot Number | 9-5501-H-03 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER |
Manufacturer Address | DEERFIELD IL |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-25 |