INTRALASE PI-RET

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-25 for INTRALASE PI-RET manufactured by Johnson & Johnson Surgical Vision, Inc..

Event Text Entries

[185091104] (b)(4). The patient interface (pi) suction ring may lose suction during a procedure. Label copy states corneal fixation vacuum loss can occur. There are several factors that may contribute to suction issues such as doctor? S technique in applying the suction ring to the cornea, doctor? S technique in squeezing the pi clip to secure the suction ring to the pi cone and patient anatomy affecting the interface between the patient? S cornea and the suction ring. A manufacturer record review related to the device including device history record will be performed. Upon completion of this review, if there is any further relevant information obtained that changes the facts and/or conclusion, a supplemental medwatch will be filed. All pertinent information available to johnson & johnson surgical vision, inc. Has been submitted.
Patient Sequence No: 1, Text Type: N, H10


[185091105] The surgery center reported suction loss while laser firing. The surgery center realigned the same patient interface cone and tried again but lost suction a second time. The procedure was aborted. The patient will be brought at a later date to complete the procedure and convert to a photorefractive keratectomy (prk) after the eye is healed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006695864-2020-00201
MDR Report Key9878626
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-03-25
Date of Report2020-03-25
Date of Event2020-03-05
Report Date2005-01-01
Date Reported to FDA2005-01-01
Date Reported to Mfgr2005-01-10
Date Mfgr Received2020-03-05
Date Added to Maude2020-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSOMYATA NAGPAL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478200
Manufacturer G1JOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer Street510 COTTONWOOD DRIVE
Manufacturer CityMILPITAS CA 95035
Manufacturer CountryUS
Manufacturer Postal Code95035
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameINTRALASE
Generic NamePATIENT INTERFACE
Product CodeHNO
Date Received2020-03-25
Model NumberPI-RET
Lot Number60207828
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerJOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer Address1700 E ST ANDREW PLACE SANTA ANA CA 92705 US 92705

Device Sequence Number: 101

Product Code---
Date Received2020-03-25
Device Sequence No101
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-25