SITE-RITE 8 ULTRASOUND 9770501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-25 for SITE-RITE 8 ULTRASOUND 9770501 manufactured by Bard Access Systems.

Event Text Entries

[186722354] The information provided by bd represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd. The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records. Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation. During evaluation, the reported issue of not holding a charge was confirmed. The sr8 was received with no physical damage noted. The sr8 was powered on at 94% battery life and shut off when disconnected from the ac adapter. The device was serviced, tested, and returned to the customer. A history review of serial number (b)(4) showed one other similar complaint from this serial number. The previous complaint for this serial number ((b)(4)) evaluation is still in progress (reference: mdr number 3006260740- 2020-00577)
Patient Sequence No: 1, Text Type: N, H10


[186722355] Per biomed, not holding a charge. Per evaluation outcomes: the reported problem was confirmed. The sr8 was powered on at 94% battery life and shut off when disconnected from the ac adapter.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006260740-2020-01012
MDR Report Key9878842
Report SourceUSER FACILITY
Date Received2020-03-25
Date of Report2020-03-25
Date Mfgr Received2020-03-04
Date Added to Maude2020-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSHELLY GILBERT
Manufacturer Street605 N. 5600 W.
Manufacturer CitySALT LAKE CITY UT 84116
Manufacturer CountryUS
Manufacturer Postal84116
Manufacturer Phone8015225640
Manufacturer G1DYMAX CORP.
Manufacturer Street141 ZEHNER SCHOOL ROAD
Manufacturer CityZELIENOPLE PA 16063
Manufacturer CountryUS
Manufacturer Postal Code16063
Single Use3
Remedial ActionNO
Previous Use Code3
Removal Correction Number2243072-10/11/2019-013-C
Event Type3
Type of Report3

Device Details

Brand NameSITE-RITE 8 ULTRASOUND
Generic NameSYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Product CodeIYO
Date Received2020-03-25
Returned To Mfg2020-03-13
Model Number9770501
Catalog Number9770501
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBARD ACCESS SYSTEMS
Manufacturer Address605 N. 5600 W. SALT LAKE CITY UT 84116 US 84116


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-25

© 2026 FDA.report