GREENLIGHT MOXY FIBER OPTIC 0010-2400

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-25 for GREENLIGHT MOXY FIBER OPTIC 0010-2400 manufactured by Boston Scientific Corporation.

Event Text Entries

[184915371] The device history record (dhr) confirmed that the devices met all material, assembly and performance specifications. Analysis of the device revealed the glass cap has a circumferential fracture on the distal side of fiber/cap fusion zone at the bevel edge. The fiber proximal to fracture can rotate independently of outer flow tubing. The distal part of glass cap and metal cap were detached and not returned. The glass cap exhibits severe devitrification at output window. The outer flow tubing open end exhibits minor scratch marks. Due to the observed circumferential fracture at distal side, the potential for forward firing may exist. Based on the observed circumferential fracture at distal side and metal cap detachment, an evaluation conclusion code of design inadequate for purpose of the device was assigned to this investigation. A temperature higher or close to epoxy degradation temperature near the laser beam output window may be a major impacting factor leading to epoxy failure and subsequent fiber breakage. Tissue adhesion from constant and heavy tissue contact could be the major cause resulting in the observed defects.
Patient Sequence No: 1, Text Type: N, H10


[184915372] It was reported that this complaint came in to san jose cis with no complaint information provided. Analysis of the returned device identify circumferential fracture on the distal side. This finding may potentially result in forward firing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2937094-2020-00267
MDR Report Key9879310
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-25
Date of Report2020-03-25
Date of Event2020-03-04
Date Mfgr Received2020-03-11
Device Manufacturer Date2019-03-14
Date Added to Maude2020-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactALYSON HARRIS
Manufacturer Street150 BAYTECH DRIVE
Manufacturer CitySAN JOSE CA 95134
Manufacturer CountryUS
Manufacturer Postal95134
Manufacturer Phone4089353452
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street150 BAYTECH DR.
Manufacturer CitySAN JOSE CA 95134
Manufacturer CountryUS
Manufacturer Postal Code95134
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGREENLIGHT MOXY FIBER OPTIC
Generic NamePOWERED LASER SURGICAL INSTRUMENT
Product CodeGEX
Date Received2020-03-25
Returned To Mfg2020-03-04
Model Number0010-2400
Catalog Number0010-2400
Lot Number0023487673
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address150 BAYTECH DRIVE SAN JOSE CA 95134 US 95134


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.