COYOTE 24698

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-03-25 for COYOTE 24698 manufactured by Boston Scientific Corporation.

Event Text Entries

[185020010] (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[185020011] It was reported that balloon rupture occurred. The 100% stenosed target lesion was located in the moderately tortuous and severely calcified vessel below the knee. A 3. 0mm x 150mm x 150cm, mr coyote balloon catheter was advanced for dilatation. However, during the first inflation at 8 atmospheres, the balloon ruptured. Subsequently, another 3. 0mm x 150mm x 150cm, mr coyote balloon catheter was used but it also ruptured during the first inflation 8 atmospheres. The procedure was completed using a different device. There were no patient complications nor injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-03416
MDR Report Key9879326
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-03-25
Date of Report2020-03-25
Date of Event2020-03-03
Date Mfgr Received2020-03-03
Device Manufacturer Date2019-11-12
Date Added to Maude2020-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal Code55311
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOYOTE
Generic NameCATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Product CodeLIT
Date Received2020-03-25
Model Number24698
Catalog Number24698
Lot Number0024753381
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-25

© 2026 FDA.report