MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-25 for DUAL CHAMBER TEMPORARY PACEMAKER 5392 manufactured by Plexus Manufacturing Sdn. Bhd.
[184996326]
Product analysis: manufacturer's analysis indicated that the external pulse generator (epg) display was out of specification, and that the ventricular output connector was broken. It was also indicated that the case was damaged. The epg was repaired and returned to use. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[184996327]
The external pulse generator (epg) was returned for testing and calibration and subsequently tested out of specification during manufacturer's analysis. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3004593495-2020-00325 |
MDR Report Key | 9880172 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-25 |
Date of Report | 2020-03-25 |
Date of Event | 2020-02-28 |
Date Mfgr Received | 2020-03-10 |
Device Manufacturer Date | 2015-04-22 |
Date Added to Maude | 2020-03-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAULA BIXBY |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635055378 |
Manufacturer G1 | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Street | BAYAN LEPAS FREE INDUSTRIAL ZO |
Manufacturer City | BAYAN LEPAS 11900 |
Manufacturer Country | MY |
Manufacturer Postal Code | 11900 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | DUAL CHAMBER TEMPORARY PACEMAKER |
Generic Name | PULSE-GENERATOR, PACEMAKER, EXTERNAL |
Product Code | DTE |
Date Received | 2020-03-25 |
Returned To Mfg | 2020-03-06 |
Model Number | 5392 |
Catalog Number | 5392 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PLEXUS MANUFACTURING SDN. BHD |
Manufacturer Address | BAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-25 |