INQUIRY? STEERABLE DIAGNOSTIC CATHETER IBI-81174

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-25 for INQUIRY? STEERABLE DIAGNOSTIC CATHETER IBI-81174 manufactured by St. Jude Medical, Costa Rica Ltda.

Event Text Entries

[186574153] One decapolar, inquiry steerable diagnostic catheter was received for evaluation. The device was returned due to a display issue. Electrode 4 read as an open circuit. A biological foreign material was noted covering electrode 4, interfering with the conductivity of the electrode and consistent with the open circuit detected and the reported display issue. Electrodes 1-3 and 5-10 met specifications of acceptable resistance values. The device history record was reviewed to ensure that each manufacturing and inspection operation was performed. The cause of the electrical issue is due to the biological foreign material covering the electrode.
Patient Sequence No: 1, Text Type: N, H10


[186574154] This report is to advise of an event observed during analysis, confirming tissue on the returned device.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008452825-2020-00164
MDR Report Key9880390
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-25
Date of Report2020-03-25
Date of Event2020-01-07
Date Mfgr Received2020-03-17
Device Manufacturer Date2019-10-04
Date Added to Maude2020-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL, COSTA RICA LTDA
Manufacturer StreetPARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL
Manufacturer CityALAJUELA, COSTA RICA 1897-4050
Manufacturer CountryCS
Manufacturer Postal Code1897-4050
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINQUIRY? STEERABLE DIAGNOSTIC CATHETER
Generic NameCATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Product CodeDRF
Date Received2020-03-25
Returned To Mfg2020-02-04
Model NumberIBI-81174
Catalog NumberIBI-81174
Lot Number7211538
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, COSTA RICA LTDA
Manufacturer AddressPARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL ALAJUELA, COSTA RICA 1897-4050 CS 1897-4050


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-25

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.