MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-26 for CARELINK PROGRAMMER 2090W manufactured by Medtronic, Inc..
[187211732]
Product analysis: the analysis could not confirm the customer comment of the programmer does not auto select, is slow to boot up and stops working. It was determined at analysis that the keyboard is separating at the hinge pin. The ac power cord is not hospital grade. All found defective parts were replaced and all issues were resolved. The programmer passed final functional and system tests. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[187211733]
It was reported the programmer is slow to boot up, and the programmer stops working. The programmer does not auto select and is unable to select the device. The programmer was returned to service and repair. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2182208-2020-00585 |
MDR Report Key | 9881968 |
Report Source | COMPANY REPRESENTATIVE |
Date Received | 2020-03-26 |
Date of Report | 2020-03-26 |
Date of Event | 2020-02-27 |
Date Mfgr Received | 2020-02-27 |
Device Manufacturer Date | 2011-06-15 |
Date Added to Maude | 2020-03-26 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAULA BIXBY |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635055378 |
Manufacturer G1 | MEDTRONIC, INC. |
Manufacturer Street | 8200 CORAL SEA STREET NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal Code | 55112 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CARELINK PROGRAMMER |
Generic Name | PROGRAMMER, PACEMAKER |
Product Code | KRG |
Date Received | 2020-03-26 |
Returned To Mfg | 2020-03-03 |
Model Number | 2090W |
Catalog Number | 2090W |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC, INC. |
Manufacturer Address | 8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-26 |