CARELINK PROGRAMMER 2090W

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-26 for CARELINK PROGRAMMER 2090W manufactured by Medtronic, Inc..

Event Text Entries

[187211732] Product analysis: the analysis could not confirm the customer comment of the programmer does not auto select, is slow to boot up and stops working. It was determined at analysis that the keyboard is separating at the hinge pin. The ac power cord is not hospital grade. All found defective parts were replaced and all issues were resolved. The programmer passed final functional and system tests. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[187211733] It was reported the programmer is slow to boot up, and the programmer stops working. The programmer does not auto select and is unable to select the device. The programmer was returned to service and repair. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182208-2020-00585
MDR Report Key9881968
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-02-27
Date Mfgr Received2020-02-27
Device Manufacturer Date2011-06-15
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC, INC.
Manufacturer Street8200 CORAL SEA STREET NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal Code55112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK PROGRAMMER
Generic NamePROGRAMMER, PACEMAKER
Product CodeKRG
Date Received2020-03-26
Returned To Mfg2020-03-03
Model Number2090W
Catalog Number2090W
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC.
Manufacturer Address8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.