MICRA MC1VR01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-26 for MICRA MC1VR01 manufactured by Medtronic Ireland.

Event Text Entries

[186517956] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[186517957] It was reported that increased thresholds and intermittent pacing were noted on the leadless implantable pulse generator (ipg) during a pre discharge device check. Dislodgement was suspected by x-ray. The device was reprogrammed and then explanted and replaced. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612164-2020-01356
MDR Report Key9882075
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-03-02
Date Mfgr Received2020-03-02
Device Manufacturer Date2019-09-12
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC IRELAND
Manufacturer StreetPARKMORE BUSINESS PARK WEST
Manufacturer CityGALWAY
Manufacturer CountryIE
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMICRA
Generic NameLEADLESS PACEMAKER
Product CodePNJ
Date Received2020-03-26
Model NumberMC1VR01
Catalog NumberMC1VR01
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC IRELAND
Manufacturer AddressPARKMORE BUSINESS PARK WEST GALWAY IE


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-03-26

© 2026 FDA.report