DI QUICK CONNECT POLYAXIAL SCREWDRIVER 279722400

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-26 for DI QUICK CONNECT POLYAXIAL SCREWDRIVER 279722400 manufactured by Depuy Spine Inc.

Event Text Entries

[186415270] The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[186415271] Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2020 during a posterior spinal fusion (psf) procedure to treat a lumbar spinal stenosis while the screw (179722945) was being inserted into the sturdy bone, the tip of the screwdriver (2797224009) broke. The fragmented tip and the screw were removed and the procedure was completed with replacements with less than a 30-minute surgical delay. Concomitant device: exp di poly screw 9x45mm ti (part#: 179722945, lot#: rl267374, quantity: 1). This report is for one (1) di quick connect polyaxial screwdriver. This is report 1 of 1 for (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1526439-2020-00768
MDR Report Key9882684
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-26
Date of Report2020-02-28
Date of Event2020-02-28
Date Mfgr Received2020-02-28
Device Manufacturer Date2017-07-18
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone6103142063
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDI QUICK CONNECT POLYAXIAL SCREWDRIVER
Generic NameSCREWDRIVERS
Product CodeHXX
Date Received2020-03-26
Returned To Mfg2020-03-23
Model Number279722400
Catalog Number279722400
Lot NumberMI60389
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SPINE INC
Manufacturer Address325 PARAMOUNT DRIVE RAYNHAM MA 02767 US 02767


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2026 FDA.report