QUICKFLEX MICRO LV LEAD 1258T/86 1258T-86

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-26 for QUICKFLEX MICRO LV LEAD 1258T/86 1258T-86 manufactured by St. Jude Medical, Inc.(crm-sunnyvale).

Event Text Entries

[185050342] The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10


[185050343] Related manufacturer reference number: 2938836-2020-02201, 2938836-2020-02202, 2938836-2020-02203. It was reported that the device failed to convert the patient ventricular fibrillation (vf) episodes. Several shocks were delivered but did not convert the rhythm. Upon interrogation, the rhythm could be converted by its own since the most recent egms showed that the patient was no longer in vf. Undersensing was observed possibly due to small waveforms. Noise and sudden increase in defibrillation and pacing lead impedance were also noted on the lead. The patient expired on (b)(6) 2020.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2938836-2020-02204
MDR Report Key9883718
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-03-07
Date Mfgr Received2020-03-09
Device Manufacturer Date2011-05-05
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELIZABETH BOLTZ
Manufacturer Street15900 VALLEY VIEW COURT
Manufacturer CitySYLMAR CA 91342
Manufacturer CountryUS
Manufacturer Postal91342
Manufacturer G1ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Manufacturer Street645 ALMANOR AVENUE
Manufacturer CitySUNNYVALE CA 94085
Manufacturer CountryUS
Manufacturer Postal Code94085
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameQUICKFLEX MICRO LV LEAD
Generic NamePERMANENT PACEMAKER ELECTRODE
Product CodeOJX
Date Received2020-03-26
Model Number1258T/86
Catalog Number1258T-86
Lot Number3393647
Device Expiration Date2014-05-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Manufacturer Address645 ALMANOR AVENUE SUNNYVALE CA 94085 US 94085


Patients

Patient NumberTreatmentOutcomeDate
129388361. Death 2020-03-26