ATTUNE CONV FB PS TB TRL SZ7 2545-00-977 254500977

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-26 for ATTUNE CONV FB PS TB TRL SZ7 2545-00-977 254500977 manufactured by Depuy Ireland - 9616671.

Event Text Entries

[185050573] Product complaint #: (b)(4). Investigation summary: examination of the returned device confirmed the reported damage. The noted damage is consistent with material overload through the use of excessive force and the investigation did not establish a need for corrective action. Depuy synthes considers the investigation closed at this time. Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[185050574] It was reported that the trial had broke while trying to remove.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1818910-2020-09317
MDR Report Key9883767
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-26
Date of Report2020-03-10
Date of Event2020-03-10
Date Mfgr Received2020-03-10
Device Manufacturer Date2013-10-14
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street700 ORTHOPAEDIC DR.
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal465810988
Manufacturer Phone6107428552
Manufacturer G1DEPUY ORTHOPAEDICS, INC. 1818910
Manufacturer Street700 ORTHOPAEDIC DR.
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal Code465810988
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameATTUNE CONV FB PS TB TRL SZ7
Generic NameATTUNE INSTRUMENTS : TIBIAL TRIALS
Product CodeHWT
Date Received2020-03-26
Returned To Mfg2020-03-18
Model Number2545-00-977
Catalog Number254500977
Lot NumberBFA0P7S
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDEPUY IRELAND - 9616671
Manufacturer AddressLOUGHBEG RINGASKIDDY CO. CORK EI


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26