STEERABLE GUIDE CATHETER SGC0301

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-26 for STEERABLE GUIDE CATHETER SGC0301 manufactured by Abbott Vascular.

Event Text Entries

[185059176] The device was not returned for analysis. A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot. Additionally, a review of the complaint history identified no indication of a lot-specific product quality issue. A definitive cause for the reported sgc leak (loss of fluid column during sgc preparation) could not be determined. There is no indication of a product quality issue with respect to manufacture, design or labeling. The clip delivery system referenced is filed under a separate medwatch report number.
Patient Sequence No: 1, Text Type: N, H10


[185059177] This is being filed to report the leak in the sgc. It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4+. During preparation, the steerable guide catheter (sgc) failed to hold column therefore it was replaced with a new sgc. The clip delivery system (cds) was advanced to the mitral valve however the clips arms rotated without manipulation of the cds and became caught in the chordae. The clip was able to be freed without issue and implanted, reducing mr to 2. There was no clinically significant delay in the procedure and no adverse patient effects. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024168-2020-02937
MDR Report Key9883846
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-03-05
Date Mfgr Received2020-03-05
Device Manufacturer Date2019-04-29
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY BELL
Manufacturer Street26531 YNEZ RD.
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal925914628
Manufacturer Phone9519143996
Manufacturer G1ABBOTT VASCULAR, REG # 3005070406
Manufacturer Street3885 BOHANNON DRIVE
Manufacturer CityMENLO PARK CA 94025
Manufacturer CountryUS
Manufacturer Postal Code94025
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTEERABLE GUIDE CATHETER
Generic NameVALVE REPAIR
Product CodeDRA
Date Received2020-03-26
Model NumberSGC0301
Catalog NumberSGC0301
Lot Number90429U226
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT VASCULAR
Manufacturer Address26531 YNEZ RD. TEMECULA CA 925914628 US 925914628


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.