MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-26 for STEERABLE GUIDE CATHETER SGC0301 manufactured by Abbott Vascular.
[185059176]
The device was not returned for analysis. A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot. Additionally, a review of the complaint history identified no indication of a lot-specific product quality issue. A definitive cause for the reported sgc leak (loss of fluid column during sgc preparation) could not be determined. There is no indication of a product quality issue with respect to manufacture, design or labeling. The clip delivery system referenced is filed under a separate medwatch report number.
Patient Sequence No: 1, Text Type: N, H10
[185059177]
This is being filed to report the leak in the sgc. It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4+. During preparation, the steerable guide catheter (sgc) failed to hold column therefore it was replaced with a new sgc. The clip delivery system (cds) was advanced to the mitral valve however the clips arms rotated without manipulation of the cds and became caught in the chordae. The clip was able to be freed without issue and implanted, reducing mr to 2. There was no clinically significant delay in the procedure and no adverse patient effects. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2024168-2020-02937 |
MDR Report Key | 9883846 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-26 |
Date of Report | 2020-03-26 |
Date of Event | 2020-03-05 |
Date Mfgr Received | 2020-03-05 |
Device Manufacturer Date | 2019-04-29 |
Date Added to Maude | 2020-03-26 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. LINDSEY BELL |
Manufacturer Street | 26531 YNEZ RD. |
Manufacturer City | TEMECULA CA 925914628 |
Manufacturer Country | US |
Manufacturer Postal | 925914628 |
Manufacturer Phone | 9519143996 |
Manufacturer G1 | ABBOTT VASCULAR, REG # 3005070406 |
Manufacturer Street | 3885 BOHANNON DRIVE |
Manufacturer City | MENLO PARK CA 94025 |
Manufacturer Country | US |
Manufacturer Postal Code | 94025 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | STEERABLE GUIDE CATHETER |
Generic Name | VALVE REPAIR |
Product Code | DRA |
Date Received | 2020-03-26 |
Model Number | SGC0301 |
Catalog Number | SGC0301 |
Lot Number | 90429U226 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT VASCULAR |
Manufacturer Address | 26531 YNEZ RD. TEMECULA CA 925914628 US 925914628 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-26 |