ARCHITECT I2000SR ANALYZER 3M74-02 03M74-88

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-03-26 for ARCHITECT I2000SR ANALYZER 3M74-02 03M74-88 manufactured by Abbott Laboratories (irving Ia/cc).

Event Text Entries

[185066645] During the subsequent site visit by the field service engineer (fse) the analyzer was inspected and discovered that the architect vacuum pump was the likely cause due to the damaged vacuum pump capacitor. The vacuum pump was replaced by the fse which resolved the issue. A review of the service history for the architect isr01183 analyzer revealed no additional issues were reported. A review of tracking and trending did not identify any trends for the architect i2000sr or the architect vacuum pump. Labeling was reviewed and contains adequate information on troubleshooting, removal and replacement of the part. The damage observed was limited to a small area of the vacuum pump and the damage did not spread to other parts of the instrument. Based on the investigation, no product deficiency was identified.
Patient Sequence No: 1, Text Type: N, H10


[185066646] The customer reported smoke emanating from the architect i2000sr analyzer from the left rear side facing the front. The customer powered off the analyzer. No injuries occurred due to this issue. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3016438761-2020-00013
MDR Report Key9884061
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-03-09
Date Mfgr Received2020-03-09
Device Manufacturer Date2003-09-18
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTIAN LEE
Manufacturer Street100 ABBOTT PARK ROAD DEPT 09B9, BLDG CP01-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224668-294
Manufacturer G1ABBOTT LABORATORIES (IRVING IA/CC)
Manufacturer Street1915 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT I2000SR ANALYZER
Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Product CodeJJE
Date Received2020-03-26
Model Number3M74-02
Catalog Number03M74-88
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES (IRVING IA/CC)
Manufacturer Address1915 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.