MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-03-26 for ALINITY S ANTI-HCV REAGENT KIT 06P04-55 manufactured by Abbott Gmbh.
[185107596]
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. This report is being filed on an international product, list number 06p04-55 that has a similar product distributed in the us, list number 06p04-60. Complete information section a, 1. Patient identifier = 102720001689. There was no additional patient information provided due to privacy issues.
Patient Sequence No: 1, Text Type: N, H10
[185107597]
The customer reported false reactive alinity s anti-hcv results on one donor. The results provided were: on 04mar2020 sid102720001689 =0. 97s/co (<1. 00s/co = nonreactive) on alinitys serial number (sn) (b)(4). The customer did not like the result being close to the cutoff, so on 05mar2020 the sample was retested on alinity s sn (b)(4) and the results =1. 20s/co (>1. 00s/co = reactive) / retest = 0. 98 and 1. 10s/co. The sample was sent and tested by nat and realtime pcr with both being negative. There was no reported impact to donor management.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3002809144-2020-00218 |
MDR Report Key | 9884212 |
Report Source | FOREIGN,HEALTH PROFESSIONAL |
Date Received | 2020-03-26 |
Date of Report | 2020-03-26 |
Date of Event | 2020-03-05 |
Date Mfgr Received | 2020-03-06 |
Date Added to Maude | 2020-03-26 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | CHRISTIAN LEE |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT 09B9, BLDG CP01-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 224668-294 |
Manufacturer G1 | ABBOTT GMBH |
Manufacturer Street | MAX-PLANCK-RING 2 |
Manufacturer City | WIESBADEN 65205 |
Manufacturer Country | GM |
Manufacturer Postal Code | 65205 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ALINITY S ANTI-HCV REAGENT KIT |
Generic Name | ANTI-HCV |
Product Code | MZO |
Date Received | 2020-03-26 |
Catalog Number | 06P04-55 |
Lot Number | 09221BE00 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT GMBH |
Manufacturer Address | MAX-PLANCK-RING 2 WIESBADEN 65205 GM 65205 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-26 |