ALINITY S ANTI-HCV REAGENT KIT 06P04-55

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-03-26 for ALINITY S ANTI-HCV REAGENT KIT 06P04-55 manufactured by Abbott Gmbh.

Event Text Entries

[185107596] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. This report is being filed on an international product, list number 06p04-55 that has a similar product distributed in the us, list number 06p04-60. Complete information section a, 1. Patient identifier = 102720001689. There was no additional patient information provided due to privacy issues.
Patient Sequence No: 1, Text Type: N, H10


[185107597] The customer reported false reactive alinity s anti-hcv results on one donor. The results provided were: on 04mar2020 sid102720001689 =0. 97s/co (<1. 00s/co = nonreactive) on alinitys serial number (sn) (b)(4). The customer did not like the result being close to the cutoff, so on 05mar2020 the sample was retested on alinity s sn (b)(4) and the results =1. 20s/co (>1. 00s/co = reactive) / retest = 0. 98 and 1. 10s/co. The sample was sent and tested by nat and realtime pcr with both being negative. There was no reported impact to donor management.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002809144-2020-00218
MDR Report Key9884212
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-03-05
Date Mfgr Received2020-03-06
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTIAN LEE
Manufacturer Street100 ABBOTT PARK ROAD DEPT 09B9, BLDG CP01-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224668-294
Manufacturer G1ABBOTT GMBH
Manufacturer StreetMAX-PLANCK-RING 2
Manufacturer CityWIESBADEN 65205
Manufacturer CountryGM
Manufacturer Postal Code65205
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameALINITY S ANTI-HCV REAGENT KIT
Generic NameANTI-HCV
Product CodeMZO
Date Received2020-03-26
Catalog Number06P04-55
Lot Number09221BE00
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT GMBH
Manufacturer AddressMAX-PLANCK-RING 2 WIESBADEN 65205 GM 65205


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26