MINICAP TRANSFER SET 5C4482

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-26 for MINICAP TRANSFER SET 5C4482 manufactured by Baxter Healthcare Corporation.

Event Text Entries

[185070220] The device was received for evaluation. Visual inspection was performed which identified a broken occluder feet. The reported condition was verified as a broken occluder feet would result in a leak. The cause of the broken occluder feet could not be determined. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[185070221] It was reported that a minicap transfer set had fluid flow with the twist clamp closed. This occurred during use of the device for peritoneal dialysis therapy. The transfer set was in use two (2) days. There was no patient injury however prophylactic anti-inflammatory medication was administered. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2020-01739
MDR Report Key9884239
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-03-02
Date Mfgr Received2020-03-03
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - MOUNTAIN HOME
Manufacturer Street1900 N HIGHWAY 201
Manufacturer CityMOUNTAIN HOME AR 72653
Manufacturer CountryUS
Manufacturer Postal Code72653
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMINICAP TRANSFER SET
Generic NameSET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Product CodeKDJ
Date Received2020-03-26
Returned To Mfg2020-03-16
Model NumberNA
Catalog Number5C4482
Lot NumberH19A08042
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE CORPORATION
Manufacturer AddressDEERFIELD IL


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.