MITRACLIP? SYSTEM CLIP DELIVERY SYSTEM CDS0601-XTR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-26 for MITRACLIP? SYSTEM CLIP DELIVERY SYSTEM CDS0601-XTR manufactured by Abbott Vascular.

Event Text Entries

[185093114] The device was not returned for analysis. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot. Additionally, a review of the complaint history identified no lot specific product issue. All available information was investigated and a definitive cause for the reported slda could not be determined in this complaint; however, the recurrent mr was due to slda. The reported patient effect of mitral regurgitation as listed in the mitraclip system instructions for use (ifu), is a known possible complication associated with mitraclip procedures. There is no indication of product quality issue with respect to manufacture, design or labeling.
Patient Sequence No: 1, Text Type: N, H10


[185093115] This is being filed to report that the clip detached from one leaflet. It was reported that the mitraclip procedure was performed to treat degenerative mitral regurgitation (mr) with grade 4. The clip delivery system (cds) was advanced to the mitral valve and the clip was deployed successfully, reducing mr to 2. However, when the cds was removed, it was noted that the clip detached from the posterior leaflet and remained attached to the anterior leaflet, single leaflet device attachment (slda) and mr increased to 4. The patient is stable. In two weeks, the patient will be evaluated for surgical repair. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024168-2020-02952
MDR Report Key9884267
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-03-06
Date Mfgr Received2020-03-06
Device Manufacturer Date2019-09-19
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY BELL
Manufacturer Street26531 YNEZ RD.
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal925914628
Manufacturer Phone9519143996
Manufacturer G1ABBOTT VASCULAR, REG # 3005070406
Manufacturer Street3885 BOHANNON DRIVE
Manufacturer CityMENLO PARK CA 94025
Manufacturer CountryUS
Manufacturer Postal Code94025
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMITRACLIP? SYSTEM CLIP DELIVERY SYSTEM
Generic NameVALVE REPAIR
Product CodeNKM
Date Received2020-03-26
Model NumberCDS0601-XTR
Catalog NumberCDS0601-XTR
Lot Number90918U121
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT VASCULAR
Manufacturer Address26531 YNEZ RD. TEMECULA CA 925914628 US 925914628


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2024 FDA.report