SUMMIT STANDARD IMP. INSERTER 2570-05-100 257005100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-26 for SUMMIT STANDARD IMP. INSERTER 2570-05-100 257005100 manufactured by Depuy Orthopaedics Inc Us.

Event Text Entries

[185098926] Product complaint #: (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[185098927] The dr was preparing to insert a corail stem into a patient, and upon inserting the stem impactor into the stem implant, the impactor would not seat down all of the way. He was able to use the instrument to insert the implant and finish the case. After the case, upon inspection, the sides of the impactor were flattened over from wear, and this was what kept the instrument from being able to be fully inseerted into the implant.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1818910-2020-09352
MDR Report Key9884524
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-26
Date of Report2020-03-13
Date of Event2020-03-13
Date Mfgr Received2020-03-13
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street700 ORTHOPAEDIC DRIVE
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal465810988
Manufacturer Phone6107428552
Manufacturer G1DEPUY ORTHOPAEDICS, INC. 1818910
Manufacturer Street700 ORTHOPAEDIC DR.
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal Code465810988
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSUMMIT STANDARD IMP. INSERTER
Generic NameHIP INSTRUMENTS : IMPACTORS
Product CodeHWR
Date Received2020-03-26
Model Number2570-05-100
Catalog Number257005100
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY ORTHOPAEDICS INC US
Manufacturer Address700 ORTHOPAEDIC DRIVE WARSAW IN 465810988 US 465810988


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2026 FDA.report