NC QUANTUM APEX 6531

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-03-26 for NC QUANTUM APEX 6531 manufactured by Boston Scientific Corporation.

Event Text Entries

[185212722] Age at time of event: 18 years or older. Device evaluated by mfr: returned product consisted of an nc quantum apex balloon catheter. The shaft, hypotube, tip and balloon were microscopically and visually examined. There was contrast in the inflation lumen and blood in the guidewire lumen. The balloon was loosely folded. Inspection of the device revealed numerous kinks throughout the hypotube and there was a complete hypotube separation 43. 5cm distal of the strain relief. The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation. Also, the tip was damaged.
Patient Sequence No: 1, Text Type: N, H10


[185212723] Reportable based on device analysis completed on 03/05/2020. It was reported that the shaft was damaged. A 12mm x 2. 50mm nc quantum apex was selected however, during unpacking, it was noted that the shaft was damaged. The procedure was completed with a different device. No patient complications were reported and patient status was stable. However, device analysis revealed a shaft break.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-03500
MDR Report Key9884870
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2019-12-27
Date Mfgr Received2020-03-05
Device Manufacturer Date2019-04-08
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal Code55311
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNC QUANTUM APEX
Generic NameCATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Product CodeLOX
Date Received2020-03-26
Returned To Mfg2020-02-15
Model Number6531
Catalog Number6531
Lot Number0023607075
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.