PASSPORT 2 MONITOR 0998-00-0900-5014A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-26 for PASSPORT 2 MONITOR 0998-00-0900-5014A manufactured by Shenzhen Mindray Bio-medical Electronics Co., Ltd.

Event Text Entries

[185129536] The user facility requested an evaluation of the equipment; no malfunction of the equipment was claimed. The results of the evaluation determined the monitor alarm volume was set to its minimum level. It was also observed that the heart rate numerics were displayed with a four second delay, however, high and low heart rate alarms were confirmed to have been called visually and audibly. Movement of the esis trunk cable caused ecg waveform artifact resulting in the delay of the heart rate numerics. The cable was replaced and there was no further delay with the display of heart rate numerics.
Patient Sequence No: 1, Text Type: N, H10


[185129537] It was reported that a patient monitored on a passport 2 monitor and gas module was having surgery and one minute into the procedure the patient's heart stopped; the patient later expired.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3009156722-2020-00005
MDR Report Key9885701
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-26
Date of Report2020-02-28
Date of Event2020-02-21
Date Mfgr Received2020-02-28
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS SERENA CHEN
Manufacturer Street1203 NANHUAN AVENUE GUANGMING DISTRICT
Manufacturer CitySHENZHEN, SHENZHEN 518106
Manufacturer CountryCH
Manufacturer Postal518106
Manufacturer G1SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD
Manufacturer StreetBUILDINGS 9-13 BAIWANGXIN HIGH-TECH INDUSTRIAL PARK
Manufacturer CityBAIMANG, XILI TOWN, NANSHAN, SHENZEN 518108
Manufacturer CountryCH
Manufacturer Postal Code518108
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePASSPORT 2 MONITOR
Generic NamePASSPORT 2 PATIENT MONITOR
Product CodeDSI
Date Received2020-03-26
Catalog Number0998-00-0900-5014A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD
Manufacturer AddressBUILDINGS 9-13 BAIWANGXIN HIGH-TECH INDUSTRIAL PARK BAIMANG, XILI TOWN, NANSHAN, SHENZEN 518108 CH 518108


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2020-03-26

© 2026 FDA.report