CARDIOSAVE HYBRID TYPE B PLUG N/A 0998-00-0800-53

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-26 for CARDIOSAVE HYBRID TYPE B PLUG N/A 0998-00-0800-53 manufactured by Datascope Corp. - Mahwah.

Event Text Entries

[185999047] The production device history record (dhr) for this iabp unit is not required to be reviewed per getinge standard operating procedure since the device manufacture date is greater than one year from the event date. (b)(6). The customer has not requested getinge service at this time. Good faith efforts (gfe) attempts are being made to the customer to obtain additional information on this complaint event. If the customer responds to our gfes with pertinent information, then a supplemental report will be sent. Not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[185999049] It was reported that on start-up, the cardiosave intra-aortic balloon pump (iabp) alarmed for "autofill failure. " the iabp would autofill after disconnecting the helium tubing and initiating 3 consecutive autofills. The iabp performed as expected when reconnected to the intra-aortic balloon (iab). There was no known harm or injury to the patient and no adverse event was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2249723-2020-00505
MDR Report Key9886304
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-26
Date of Report2020-03-26
Date of Event2020-03-03
Date Mfgr Received2020-03-03
Device Manufacturer Date2014-11-01
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street1300 MACARTHUR BLVD
Manufacturer CityMAHWAH NJ 07430
Manufacturer CountryUS
Manufacturer Postal07430
Manufacturer G1DATASCOPE CORP. - MAHWAH
Manufacturer Street1300 MACARTHUR BLVD
Manufacturer CityMAHWAH NJ 07430
Manufacturer CountryUS
Manufacturer Postal Code07430
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARDIOSAVE HYBRID TYPE B PLUG
Generic NameSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Product CodeDSP
Date Received2020-03-26
Model NumberN/A
Catalog Number0998-00-0800-53
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDATASCOPE CORP. - MAHWAH
Manufacturer Address1300 MACARTHUR BLVD MAHWAH NJ 07430 US 07430


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.