INTERSTIM II 3058

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-26 for INTERSTIM II 3058 manufactured by Medtronic Puerto Rico Operations Co..

Event Text Entries

[186133403] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[186133404] On 2020-mar-19 (b)(4) (hcp): information was received from a healthcare provider (hcp) regarding a patient with an implantable neurostimulator (ins) for urinary dysfunction/sacral nerve stimulation and gastrointestinal/pelvic floor. It was reported by the mri technician that there was a power on reset (por) message when they attempted to turn the patient? S stimulation off for an mri. The caller mentioned that they had spoken with a manufacturer representative (rep) that day but the caller indicated that the rep had been unable to meet with the patient due to covid-19. The caller noted that they (or the patient) did think that the patient had been around any emi recently. There were no symptoms reported and the caller was redirected to follow up with the managing physician as the por meant that the stimulation was off but the message could only be cleared with a clinician programmer. There were no further complications reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004209178-2020-06232
MDR Report Key9887195
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-26
Date of Report2020-03-26
Date Mfgr Received2020-03-19
Device Manufacturer Date2017-04-04
Date Added to Maude2020-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLISA WOODWARD CLARK
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635263920
Manufacturer G1MEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer StreetROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK
Manufacturer CityJUNCOS PR 00777
Manufacturer CountryUS
Manufacturer Postal Code00777
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINTERSTIM II
Generic NameSTIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Product CodeEZW
Date Received2020-03-26
Model Number3058
Catalog Number3058
Device Expiration Date2018-09-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer AddressROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK JUNCOS PR 00777 US 00777


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.